8.9 Set TDS Factor – P7.0
The alpha TDS 200 Controller allows the TDS factor to be selected, depending on the
application.
.
P.0
d
t
s
0
0
P7.1: Viewing Temperature electrode data
1. Key in the password “022” as per procedure in Section 5.
2. Press key until screen displays P7.0 and TDS. Press ENTER.
3. Screen will scroll P7.1, ‘FACT.’ then TDS factor. (Default is 0.70)
4. Press the or keys to change TDS factor (0.40 to 1.0). Press ENTER.
5. Press and keys together (ESCAPE) to return to P7.0.
8.10 Controller Reset – P8.0
Resets the controller to factory default values.
TDS range remains unchanged based on last selected range
All calibration data and TDS factor are reset
SP 1 is reset to 10 % of full scale and LOW
SP 2 is reset to 90 % of full scale and HIGH
Hysteresis for both set points is reset to 5 % of full scale
Temperature compensation mode remains unchanged
In MTC mode, temperature value is reset to 25
o
C
Temperature Coefficient is reset to 2.1%
Normalization Temperature is reset to 25.0
o
C
NOTE: ATC temperature offset is NOT reset.