Instruction Manual pH 5+, pH 6+, Ion 6+
9
Resetting User Calibrated Values
The meter can be completely reset to factory default values, or partially reset for
calibration values only using the procedure below.
Reset Type Description
No ()
No reset is performed; meter returns to measurement
Calibration ()
Reset of either pH, mV, Ion, or Temperature calibration
depending on the selected mode*
Factory ()
Reset of all calibration values and user settings to factory
default settings
*For calibration reset, measure in the mode that you want to reset prior to step 1 below.
1. Power off the meter. Press and hold
then press . If successful,” ” will
blink on the display. Release both keys.
2. Press
to enter the reset menu.
3. Press
to toggle between No Reset (), Calibration Reset (), or Factory
Reset (). Press
to cancel.
4. Press
to confirm the selected reset type. The meter will automatically begin
measurement mode.