Instruction Manual αlpha-pH2000D
- 10 -
3.2 Connection Diagram
Caution: Ensure electrical mains are disconnected before proceeding.
Connections for wall mounting version
1
L
2
N
Current
15
Current
13
OP2
14
16
OP1
17
3
4
5
6
22
HOLD
18 19 2120 23 24 25
7
8 9
10
11
2926 2827 30 31 32 33
12
-5
V
TE
MP
SENS TEMP
pH
RE
F
pH
ME
AS
Fuse: 250V 315 mA,
Time Delay
L
1
12
24
Cur re nt
N
OP2
13
Cur re nt
14
2
15
16
OP1
17
3
21
HOLD
18 19 20
4
22 23
28
5
25
6 7
26 27
8
-5V
TEMP
31
TEMP
SENS
29
9
30
10
pH
REF
11
32 33
pH
ME
AS
250V, 315 mA,
Time Delay
1. AC mains live wire 18. 12V output Power supply
2. AC mains neutral wire 19. Hold Function
3. AC mains protective earth wire 20. Hold Function
4. Relay A (SP 1) 21. Earth ground
5. Relay A (SP 1) 22. no connection
6. Relay B (SP 2) 23. no connection
7. Relay B (SP 2) 24. no connection
8. Wash relay 25. no connection
9. Wash relay 26. Shield (drain wire)
10. Alarm relay (NC) 27. -5V to differential pH sensor (White wire)
11. Alarm relay common 28. Temperature ground (black wire)
12. Alarm relay (NO) 29. Temperature Sense (Yellow wire)
13. 4-20 mA temperature output, +ve terminal 30. Temperature Input (Yellow wire)
14. 4-20 mA temperature output, -ve terminal (Internally connected by jumper to terminal 29)
15. 4-20 mA pH/ORP output, +ve terminal 31. No connection
16. 4-20 mA pH/ORP output, -ve terminal 32. pH input, Reference (Green wire)
17. 12V output ground 33. pH input, Measure (Red wire)
IMPORTANT: The Alarm relay functions as an “Active Low” device i.e. it switches OFF
under Alarm condition. Therefore the Alarm display device should be connected to the
‘NC’ contacts of the relay (10 & 11).
-5V