30 Thermo Scientific Alpha pH 560
Press or key to access other setup screens or Press and
key simultaneously (escape) to return to measurement mode.
The controller settings allow you to determine the relay type and relay
status of the pH Controller.
Figure 5–24. Controller Settings
11
From pH (or ORP) measurement mode press ENT key to
enter setup mode as described in page 30. The LCD shows the
first screen of setup mode (OFS). Press or key to select
controller settings screen (Cntr).
Press ENT key to access controller settings (Cntr).
2
Select the controller type: The lower display shows ‘tyP’. The
upper display shows the last configured controller type. (‘OFF’
or ‘L.Ct’). Press or key to select the desired controller
type.
OFF:
Controller Off
Selecting ‘OFF’, the pH Controller stops functioning
as a controller by preventing relay functions. Once the
controller type is set to ‘OFF’, the setup mode does
not show you the Relay A/Relay B settings.
L.Ct:
Limit value control (Controller On)
Selecting ‘L.Ct’, the pH Controller is set to function
as a controller. When the controller type is set to
‘L.Ct’, the Relay A/Relay B settings are available in
the setup mode for you to configure relay operating
Controller
Settings