EUTECH INSTRUMENTS DO 6 PLUS Manuel d'instructions Page 31

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 48
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 30
Instruction Manual DO 6+
- 25 -
5.7 Reset to Factory Default (rSt.5)
Use this mode to reset the meter to factory default settings. There are two
levels of reset.
Calibration (A.CAL) resets calibration values only.
User (b.USR) resets all data, calibration, and other customized setup functions.
1) From the setup menu, press ▲▼ to select
“rSt.5”.
2) Press HOLD/ENTER to enter the menu.
3) Press HOLD/ENTER to enter calibration reset
“A.CAL”.
4) Use ▲▼ to select YES (reset calibration) or NO
(do not reset calibration).
5) Press HOLD/ENTER to confirm. If “YES”, the
meter returns to measurement mode after
resetting.
Note: Both % and concentration calibrations are
reset when from the % saturation setup.
However, when calibration reset occurs from
the concentration setup, only concentration is
reset.
6) If “NO”, the User Reset menu “b.USR” is
displayed.
7) Press HOLD/ENTER to enter the menu.
8) Use ▲▼ to select YES (reset user) or NO (do not reset user).
9) Press HOLD/ENTER to confirm.
10) If “YES”, the meter returns to measurement mode after resetting.
11) Press CAL to return to measurement mode.
HOLD
ENTER
HOLD
ENTER
Calibration
Reset
Vue de la page 30
1 2 ... 26 27 28 29 30 31 32 33 34 35 36 ... 47 48

Commentaires sur ces manuels

Pas de commentaire