5. SETUP FUNCTIONS
Use the setup feature to customize your instrument operation.
There are two setup menus; one menu is derived from the %
saturation mode and the other from the mg/L (or ppm) concentration
mode.
First, select the appropriate measurement mode you wish to adjust by
pressing
until the desired units are displayed. During
measurement, press and hold
for 5 seconds to enter SETUP
mode of the parameter being measured.
Press
or to change programs or change options.
Press to select the program or confirm selection.
Press to go back one level or return to measurement mode.
5.1 1.0 CAL (Calibration)
Press
to view stored calibration points.
Note: % saturation mode will list two points, while
concentration mode has only one point.
5.2 2.0 ELE (Electrode Information)
Press
to view calibration factor (CAL.F)
TIP: The factor is an indication of the probe’s efficiency. It is
the ratio of the actual mV produced by the probe to the
theoretical mV value. The ratio displays from 0.5 to 1.999.
Press
to view mV output for 100% (HI).
Press
to view mV output for 0% (LO).