Instruction Manual COND 6+, TDS 6+, SALT 6+
-5-
2.5 Battery Replacement
A “LO” annunciator in the LCD alerts you when battery power is running low. See below.
LO
S
µ
"LO" Battery Condition
Caution: Power off the meter when changing battery.
2.6 Electrode Information
Your meter includes an electrode with a BNC connector (EC-
CONSEN91B) having a nominal cell constant of k = 1.0, and a built-in
temperature sensor. The Ultem-body housing has good chemical
resistant properties. The electrode design offers fast temperature
response and reduces air entrapment, ensuring accurate, repeatable,
and stable readings.
The wettable materials of the probe include:
1. Polyetherimide (Ultem) – protective probe guard
2. Polybutylterphalate (Valox) – sensor housing
3. Stainless Steel (SS 304) – 2 steel bands
The protective probe guard can be removed temporarily for
maintenance but must be re-attached during measurement and
calibration. Erroneous results will occur while the probe guard is
removed.
Always immerse the probe beyond upper steel band for best results.
Use the fill line on the outside of the probe guard for reference.