EUTECH INSTRUMENTS SALT 6 PLUS Manuel d'instructions Page 15

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 38
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 14
Instruction Manual Salt 6
12
4.2 Preparing the Meter for Calibration
Before starting calibration, make sure you are in the correct measurement
mode.
Ensure that you use new Sodium Chloride standard solutions during calibration.
Do not reuse standard solutions as it may be contaminated and affect the
calibration and accuracy of measurements. Use fresh calibration solution each
time you calibrate your meter. Store solutions in a dry and cool environment if
possible.
Always rinse the probe with either deionized water or rinse solution before and
after each calibration/sample measurement to avoid cross-contamination. For
details please refer to Section 7 - Probe Care and Maintenance.
NOTE: These meters are factory set to a temperature
coefficient of 2.1% per °C.
The factory default value for normalization
temperature is 25 °C. If you need to normalize to a
value other than 25 °C, see Section 6.3 –
Normalization Temperature.
Vue de la page 14
1 2 ... 10 11 12 13 14 15 16 17 18 19 20 ... 37 38

Commentaires sur ces manuels

Pas de commentaire