
Instruction Manual Salt 6
21
6.2 Temperature Coefficient
The temperature coefficient is the amount of change in conductivity per degree
of temperature; it is expressed in percent per °C. Entering the exact
temperature coefficient of your solution lets you accurately compensate
temperature for almost any solution. You can adjust 0.0 to 3.0 % per °C.
Meter default is 2.1% per °C.
1) Enter the advanced setup as described in
Section 6.1.
2) Press ▲ or ▼ key until [t.Co %] appears
on the LCD. Press ENTER key.
3) Press ▲ or ▼ key to select a value
between 0.0 to 3.0.
4) Press ENTER key to select. The meter will
take you back to the menu, [t.Co %].
5) Press ▲ or ▼ key to move to the next
menu or press CAL to exit to measurement
mode.
%
%
%
HOLD
ENTER
Commentaires sur ces manuels