EUTECH INSTRUMENTS ALPHA COND 200 - REV 3 Manuel d'instructions Page 28

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 40
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 27
Operating Instructions αlpha-CON1000
23
7.7.3 Selecting the alarm time lag
This parameter group lets you select a period of time before the alarm
activates when your set point has been overshot. You can select from
0 to 1999 seconds.
1) Follow directions in 7.7.1 to enter Configuration mode.
2) Press the ENTER key. Scroll with the or keys until the upper
display shows a numerical value (in seconds) and the lower
display shows “AL.d”.
3) Press the or keys to select how long of an alarm delay (0 to 1999 seconds) you want.
4) Press the ENTER key to confirm your selection.
5) Proceed to 7.7.4, or return to Measurement mode by pressing the and keys
simultaneously (escape).
7.7.4 Selecting steady or pulse contact for the alarm relay
This parameter group lets you select whether the alarm contact will
operate as a steady contact or a fleeting (single pulse) contact. Pulse
contact closing time is 250 millisecond.
1) Follow directions in 7.7.1 to enter Configuration mode.
2) Press the ENTER key. Scroll with the or keys until the upper
display shows “Stdy” or “FLEt” and lower display shows “AL.C.”.
- AL.C = alarm contact
- StdY = steady contact
- FLEt = fleeting (single pulse) contact
3) Press the or keys to select steady or pulse contact.
4) Press the ENTER key to confirm your selection.
5) Proceed to 7.7.5, or return to Measurement mode by pressing the
and keys simultaneously (escape).
7.7.5 Wash contact (if the relay 3 is set to Wash)
1) Follow directions in 7.7.1 to enter Configuration mode.
2) Press the ENTER key until display shows “ALr” or “CLn”. Choose
“CLn” by pressing or key. Press the ENTER key.
3) Press the or key to select the wash cycle (interval 0.1 to 199.9
hours) and press the ENTER key to confirm.
4) Press the or key to select the wash duratioin (1 to 1999
seconds) and press the ENTER key to confirm.
HOLD SETUP
30
Al.
HOLD
HOLD
SETUP
SETUP
S Y
FLE
Al.C
Al.C
HOLD
HOLD
SETUP
SETUP
10.0
10
i
Vue de la page 27
1 2 ... 23 24 25 26 27 28 29 30 31 32 33 ... 39 40

Commentaires sur ces manuels