5. If calibration is successful, Controller displays DONE, blinking. Press
ENTER to revert to Conductivity measurement mode (Relays and 4-20 mA
output resume previous settings).
NOTE: To clear the ERR1 display and return to calibration mode, press /
keys together.
509
500
done
ENTER
ENTER
Notes:
You can view the calibrated Conductivity value from Setup program. See Setup
program P5.0.
Controller displays calibrated conductivity point for selected range.
If calibration is not done for the selected range, controller displays ‘- - - -‘.
If after thorough cleaning of electrode and ERR1 is displayed after an attempted
calibration, consider changing electrode.