23
8.8 Viewing Conductivity/ Temperature Electrode Data – P6.0
Program 6 has two “view only” options that let you check the electrode parameters
for diagnostic purposes.
P6.1: view revised cell constant value of electrode
P6.2: view temperature probe offset (ATC on only)
These parameters will change each time you recalibrate the controller.
P6. 0
elec
P6. 1
P6. 2
t.ofs
0. 0
1000
.
FCTR
Press and keys together (ESCAPE) at anytime, to leave the Setup mode and
return to Measurement mode.
P6.1: Viewing Conductivity electrode data
1. Key in the password “022” as per procedure in Section 5.
2. Press key until screen displays P6.0 and ELEC. Press ENTER.
3. Screen will scroll P6.1, ‘FCTR’ then cell constant factor. Each time a
calibration is performed, the value of the cell constant factor will be updated.
4. Press and keys together (ESCAPE) to return to P6.0 or press ENTER
to view Temperature electrode data.
P6.2: Viewing Temperature electrode data
5. Follow procedure from above (P6.1) until step 3 and press ENTER.
6. Screen will scroll P6.2, ‘T.OFS’ (Temperature Offset), then amount of offset
(in
o
C).
7. Press ENTER or press and keys together (ESCAPE) twice, to return to
measurement mode.