Instruction Manual PCD 650
64
8. Rinse the probe with de-ionized water. Place it in the next solution and follow
the steps 2 to 7 to calibrate other points, if ‘MULTI’ point calibration is selected.
9. Press ESC (F4) to go to measurement mode. The meter shows the TDS
reading of the solution in measurement mode.
3.9.1 TDS Calibration Report
Calibration report gives you information on calibration. It includes date & time on
which the last calibration was done and the offset.
To View TDS Report:
1. From TDS measurement mode, press left or right arrow key to navigate
to other available functions until you see REPO function in the LCD.
2. Press REPO (F2) key to go to calibration report screen (Figure 57).
3. Press PRIN (F2) to transfer the calibration report to the computer. (Refer page
37 to print data).
Figure 57 : TDS report screen
Note: The function of the keys available on the TDS cell
constant/calibration/ report screen is same as previously described in
conductivity calibration.