Instruction Manual CyberScan pH 1100/ 2100
24
6.2 Memory recall
This function recalls the previous readings stored in the memory. You can only access memory recall from the
measurement mode. Memory recall is in “Last In First Out” (LIFO) order.
There are two methods of viewing stored data in the memory recall function: manual and automatic. In manual memory
recall, the user must manually scroll through the memory points using the /MI and /MR keys. In automatic memory
recall, the meter automatically scrolls through the memory points. To choose the method of viewing stored data you wish
to use, see Meter Setup mode Program P4.3 MemRecall Setup (page 52).
6.2.1 Recalling readings in manual recall mode
1. Press the /MR key once to retrieve the last reading stored.
The display flashes Mem Recall Activated, Please wait... See
Figure 62.
The display then flashes the memory location last stored in the
lower display. See Figure 63. Then the display shows the
recorded value (pH/temp, mV or Ion/mV) in the upper display
and toggles between memory location and date in the lower
display. See Figures 64-65.
2. Use the /MR key to select the next memory location or the
/MI key to select the previous memory location.
3. Repeat step 2 to review additional stored data sets.
4. To exit Memory Recall, press EXIT and return to the
measurement mode.
6.2.2 Recalling readings in automatic recall mode
1. Press the /MR key once to retrieve the last reading stored.
The display flashes Mem Recall Activated, Please wait... See
Figure 62. The display then flashes the memory location last
stored in the lower display. See Figure 63. Then the display
shows the recorded value (pH/temp, mV or Ion/mV) in the
upper display and date and time in the lower display. See
Figure 64.
2. After 8 seconds, the meter automatically scrolls to the next to
memory location. The display flashes the next memory location
stored. See Figure 66. The display then shows the recorded
value (pH/temp, mV or Ion/mV) in the upper display and date
and time in the lower display. See Figure 67.
3. The meter will continuously scroll through all the memory points
stored indefinitely. To exit memory recall, press EXIT and return
to the measurement mode.
NOTES: If no new readings are stored after memory recall function
has been exited, the next time the memory recall function is
activated, it will resume at the memory location point where it was
last exited. If a new point is stored, the memory recall function will
begin with that point. Readings stored in memory are retained even
if the unit is turned off. To erase all readings stored in memory, see
Meter Setup Program P4.6, Mem Clear Setup (page 54).
7.545pH 25.4 C ATC
Jan 18 2002 08:50:35
°
5.695pH 25.3 C ATC
Jan 21 2002 08:52:35
°
Figure 62
Figure 63
Figure 64
Figure 65
Figure 66
Figure 67
Mem Recall Activated
Please wait. . .
Mem Loc:9
Mem Loc:8
5.695pH 25.3 C ATC
Mem Loc:9 08:52:35
°