Instruction Manual CyberScan pH 1100/ 2100
54
10.5.8 Memory clear setup mode P4.6
Clears all pH, mV and Ion (pH2100 only) memory points stored
in memory.
1. Enter meter setup mode as described in steps 1-4 in Section
10.5.1 Entering meter setup mode (page 50).
2. Use /MI key to scroll to program Mem Clear Setup: P4.6.
See Figure 181.
3. Press ENTER to enter program Mem Clear Setup: P4.6.
4. Press >/CAL key to toggle between Memory Clear Yes and No.
The selection will flash. See Figure 182.
5. Press ENTER to confirm selection.
If you select Yes, the screen flashes Memory being
Cleared, Please wait... See Figure 183. The meter then
returns to the setup mode.
If you select No, the meter returns to the setup mode.
Press /MI to continue to the next setup program, or
press /MR to go to the previous program. Press EXIT to
return to measurement mode.
10.5.9 Meter reset setup mode P4.7
Resets the meter to factory default settings:
• pH/mV/Ion* calibration data is erased
• Manual temperature compensation is set to 25.0°C
• pH/mV/Ion* measurement alarm is off
• pH/mV/Ion* calibration due alarm is off
• Stability indicator is turned on with low stability setting
• Data logging and memory recall has manual setting
• Backlight is on with 2.0 minute duration
• Memory is cleared
*Ion features on the CyberScan pH2100 meter only
1. Enter meter setup mode as described in steps 1-4 in Section
10.5.1 Entering meter setup mode (page 50).
2. Use /MI key to scroll to program Meter Reset: P4.7.
See Figure 184.
3. Press ENTER to enter program Meter Reset: P4.7.
4. Press >/CAL key to toggle between Meter Reset Yes and No.
The selection will flash. See Figure 185.
5. Press ENTER to confirm selection.
If you selected Yes, the screen flashes Meter being
Reset, Please wait...See Figure 186. The meter then
returns to measurement mode.
If you select No, the screen flashes Reset Aborted. See
Figure 187. The meter then returns to the measurement
mode.
Memory being Cleared
Please wait. . .
Memory Clear:Yes No
Select & Press ENTER
Mem Clear Setup :P4.6
Press ENT. to Select
Figure 181
Figure 182
Figure 183
Meter being Reset
Please wait. . .
Meter Reset: Yes No
Select & Press ENTER
Meter Reset :P4.7
Press ENT. to Select
Figure 184
Figure 185
Figure 186
Figure 187
Reset Aborted